Cell-free DNA · Plasma pre-analytics · Equipment configuration

A centrifuge rated at 20,600 ×g may still be unsuitable for a particular cfDNA plasma preparation step.

The answer depends on the exact rotor, the blood collection or transfer vessel, its permitted RCF, and what the assay requires at each stage—not on the centrifuge's headline specification alone.

Direct answer

In a two-stage cfDNA plasma workflow, the first centrifugation separates plasma from blood cells. After careful plasma transfer, a second, usually higher-force treatment may be specified to reduce residual cells and debris. Each stage must be checked independently against the method, rotor-specific RCF, approved vessel and processing conditions. A main-unit maximum RCF does not guarantee that every compatible rotor—or tube—can deliver the required step.

01 · FIRST SPIN Original blood collection tube

Check collection-tube geometry, additives, approved rotor and sample stability window.

02 · TRANSFER Plasma moved away from cells

Avoid disturbing the buffy coat and carrying residual blood cells into the next vessel.

03 · SECOND TREATMENT Transferred plasma vessel

Check higher RCF capability, usable fill volume, tube rating and downstream method.

Why Clear Plasma Is Not Necessarily Ready for cfDNA Analysis

Circulating cell-free DNA (cfDNA) comprises DNA fragments present in the plasma and other body fluids. Circulating tumour DNA (ctDNA) can form a low-abundance fraction of cfDNA in some samples. The analytical problem is not simply how much DNA can be recovered: it is whether the recovered material is suitable for the intended assay.

Following a routine blood separation, plasma may appear clear while still containing residual nucleated cells or cellular material. If cells are carried forward or break down during unsuitable handling or delays, they can release genomic DNA (gDNA). The additional background can affect cfDNA fragment profiles or the apparent fraction of low-abundance targets, depending on the downstream method.

A higher total DNA measurement is therefore not automatically evidence of better cfDNA preparation. When evaluating a processing method, laboratories may also need to assess fragment-size distribution, indications of high-molecular-weight DNA, extraction performance and assay-specific quality measures.

Critical distinction: A later high-speed centrifugation can help reduce residual cells and particulate material. It cannot be assumed to remove genomic DNA already released into the plasma by cells that lysed earlier in the process.

What the First and Second Centrifugation Steps Actually Do

The US National Cancer Institute (NCI) provides an evidence-based reference procedure for cfDNA blood collection and plasma processing. It differentiates whole-blood separation, careful plasma transfer and a further plasma-clearing step. Its example conditions are method-specific reference values, not a universal prescription for all liquid-biopsy assays.

Stage 1 · Whole blood

Separate plasma from blood cells

800–1,600 ×g

20 minutes in the referenced NCI procedure, at room temperature or 4°C as specified. The original blood collection tube and its permitted conditions must be checked.

Stage 2 · Transferred plasma

Reduce residual cellular material

14,000–16,000 ×g

10–20 minutes in the referenced NCI procedure, at room temperature or 4°C as specified. An appropriate filtration step is an acceptable alternative in that guidance.

Between these stages, plasma is transferred into a new, method-appropriate vessel while leaving the buffy coat behind. This physical transfer is a key part of the workflow: spinning intact whole blood at a higher force does not reproduce the intermediate separation and transfer.

Not every cfDNA protocol requires exactly two spins. Different clinical or research methods may validate a different sequence, alternative RCF, filtration or other sample-processing conditions. Never substitute the NCI reference values for an assay-specific or collection-tube-specific validated protocol.

Why One High-Speed Spin Cannot Automatically Replace Two Stages

It may be tempting to use the higher RCF from the second stage directly on the original whole-blood tube, reasoning that one stronger spin should do both jobs. That conclusion ignores a fundamental difference between the stages: the second spin acts on plasma that has already been removed from the blood-cell layer.

Before considering any single-spin alternative, a laboratory would have to evaluate the actual collection vessel's allowed RCF, the sample's cellular carryover, the possibility of cell damage, the handling sequence and the downstream assay's acceptance criteria. If a validated method expressly uses one spin, it should be assessed on its own evidence; the point is that a higher maximum RCF alone cannot prove equivalence.

Two Different Vessels Mean Two Different Compatibility Decisions

Purchasers sometimes combine their requirements into a sentence such as, “We process 10 mL blood tubes and need 16,000 ×g.” That request does not yet identify which vessel must withstand 16,000 ×g, or how much plasma is transferred into it.

Check First spin: collection tube Second spin: plasma transfer vessel
Sample Anticoagulated whole blood Separated plasma, with cellular interface avoided
Vessel identity Exact collection tube, additive/stabiliser, dimensions and cap Exact transfer tube, material, cap, dimensions and low-DNA-binding suitability if required
Rotor Verified swing-out or fixed-angle configuration and correct adapter Verified higher-force rotor with the correct vessel support
Permitted RCF Both rotor and collection-tube limits Both rotor and transfer-tube limits, including fill restrictions
Capacity unit Original blood tubes per run Plasma aliquots or transferred tubes per run
Additional conditions Processing delay, temperature and handling Timing, temperature, transfer method and downstream assay

Nominal “10 mL” capacity does not establish that any 10 mL tube fits a rotor or can be centrifuged at its maximum RCF. Tube geometry, closure, material, adapter and permissible filling are part of the approved operating combination. A rotor listing a 10 mL position is not, by itself, approval for a particular blood collection or plasma tube.

The Real Procurement Trap: Maximum Machine RCF Is Not Rotor RCF

GlanLab's YT16R illustrates why the selected rotor matters. Its main unit is rated up to 16,000 rpm and 20,600 ×g, but its published compatible rotor options have different maximum RCF values and vessel capacities.

GlanLab YT16R rotor Published capacity Max. RCF What it means for a 14,000–16,000 ×g method
G16-8 24 × 1.5/2 mL 20,600 ×g Force capacity exceeds the reference range; usable volume and transfer-vessel approval remain to be confirmed.
G16-13 6 × 10 mL 16,460 ×g Published rotor capability reaches the reference range, subject to tube compatibility and correct settings.
G16-14 12 × 10 mL 14,510 ×g Can reach the lower part of the reference range, but not a method requiring 16,000 ×g.
G16-15 12 × 15 mL 11,840 ×g Cannot meet a protocol with a minimum 14,000 ×g requirement, despite the larger nominal vessel capacity.

A specific rejection case: If the method requires at least 14,000 ×g at the actual sample position, the G16-15's published 11,840 ×g maximum is insufficient. Do not approve the configuration just because the YT16R main unit is advertised at 20,600 ×g. Conversely, even a rotor that meets RCF is not approved for a tube until the complete vessel/rotor combination is verified.

The applicable maximum operating force is constrained by every part of the system: centrifuge, exact rotor, bucket/adapter and vessel (including fill and sample-density conditions). The published figures are for the stated rotor configurations and should be reconfirmed in the final quotation.

GlanLab Configurations for the Two Processing Stages

The first stage can place greater emphasis on compatible blood collection tubes and batch handling, while the second requires confirmation of a substantially higher-force configuration. These are examples of capability matching, not product validation for a particular cfDNA assay.

GlanLab YT5AR refrigerated centrifuge for blood collection tube separation
YT5AR · Lower-force stage Up to 5,000 rpm and 4,730 ×g at main-unit level. The 5A-7 option includes a published 4 × 12 × 10 mL vacuum-tube configuration rated at up to 2,790 ×g.
GlanLab YT16R high speed refrigerated centrifuge for rotor-specific plasma processing
YT16R · Higher-force stage Up to 16,000 rpm and 20,600 ×g at main-unit level. Actual limits depend on the rotor; the G16-13, G16-14 and G16-15 have different rated maxima.

For the first stage, the YT5AR and its 5A-7 blood-tube rotor options provide a starting point for assessing large batches of original collection tubes. The listed 4 × 12 × 10 mL vacuum-tube arrangement has a published maximum of 2,790 ×g, covering the force envelope of the NCI example in principle. Actual tube identity, dimensions, adapters, cap clearance, allowed force, method temperature and processing time still require confirmation.

For a high-force second stage, the YT16R rotor specification table allows selection to begin with the real plasma vessel and required RCF. A smaller-volume rotor may provide the required force but create many aliquots; a larger-volume rotor may reduce handling while failing the force requirement. The appropriate choice depends on both constraints.

GlanLab 5A-7 swing-out rotor configuration for 4 by 12 blood collection tubes of 10 mL nominal size
5A-7 · 4 × 12 × 10 mL vacuum-tube option Published maximum RCF: 2,790 ×g for this configuration.
GlanLab G16-13 six-position 10 mL angle rotor for YT16R
G16-13 · 6 × 10 mL angle rotor Published maximum RCF: 16,460 ×g; not a blanket approval for all 10 mL tubes.

Why Sample Count and Plasma Volume Produce Different Throughput Numbers

Capacity for the first spin is normally counted in original blood tubes. The second spin is governed by transferred plasma volume, permissible working fill per tube, aliquot tracking and the number of supported rotor positions. The two stages may therefore have very different batch counts.

Illustrative workload calculation—not a YT16R throughput claim: Suppose 24 blood samples each yield 2.4 mL of plasma, and a separately verified transfer method permits 1.5 mL of usable plasma per high-force vessel. Keeping each sample separate requires 24 × ceil(2.4 ÷ 1.5) = 48 vessels. A 24-position rotor would require at least 2 complete loading cycles to process those vessels, before considering allowed balance patterns, handling and run time.

The result changes when usable filling volume, aliquot requirements or rotor choice changes. It cannot be inferred from a headline “24 × 2 mL” capacity without checking the actual allowed vessel filling and protocol.

Second-stage vessel count = Σ ceil(plasma volume per sample ÷ approved usable volume per vessel)

This calculation is particularly useful for labs comparing a higher-force microtube rotor with a lower-force larger-volume rotor. More nominal millilitres are not useful if the specified RCF cannot be reached.

Processing Delays and Collection Tubes Matter Before Either Spin

Pre-analytical quality can deteriorate before centrifugation begins. If nucleated blood cells break down while whole blood is being stored or transported, genomic DNA can enter the plasma. Subsequent centrifugation should not be assumed to reverse that release.

The NCI reference procedure identifies processing within two hours as optimal for EDTA blood, while also discussing conditional longer intervals and specific stabilising collection-tube alternatives. These are not universal limits for every collection tube or assay. Stabilising-tube requirements can differ from standard EDTA handling and must be checked against the actual method and tube instructions.

For international laboratories, this is a procurement issue as well as a procedure issue: processing sites must be able to meet the required collection-to-spin window, temperature, transport conditions and sample-transfer capacity. A centrifuge chosen solely by RCF cannot compensate for an unworkable sample logistics plan.

How to Decide Whether the Second Spin Is Necessary

A further high-force spin is included in the cited NCI reference procedure, which also accepts an appropriate filtration alternative. However, the decision for a particular assay should follow its validated processing method and evidence for the relevant sample and analytical endpoint—not a general assumption that more centrifugation always improves every cfDNA result.

When considering a change from double-spin to single-spin processing, a laboratory should evaluate relevant quality measures, including residual cellular material, high-molecular-weight DNA, fragment profiles, recovery and assay-specific performance. Matching total DNA yield alone is not enough to establish equivalence.

A Practical cfDNA Centrifuge RFQ: Specify Both Stages

A quote request that simply asks for a “16,000 ×g refrigerated centrifuge” leaves critical compatibility and capacity questions unanswered. Define the working configuration first.

RFQ field What to provide
Application and intended use cfDNA plasma preparation, assay/research type and relevant method or SOP
Stage-by-stage conditions RCF, duration, temperature and any specified braking/acceleration conditions
Original collection tube Exact tube identity, additive, nominal fill, dimensions and allowed centrifugal conditions
Plasma transfer vessel Exact model/dimensions, cap, material and approved working fill at the required RCF
Rotor and adapters Exact rotor order number, buckets, inserts and approved combinations for each stage
Workload Original samples per batch, expected plasma yield, aliquots per sample and peak arrival pattern
Workflow constraints Time from collection to processing, plasma transfer steps, number of parallel units and required continuity
Procurement details Destination country, voltage/frequency, quantities and documentation required for intended use

Laboratories should distinguish research-use suitability from any regulatory or diagnostic requirements applicable to the actual intended use. Do not infer a clinical approval from a centrifuge's general specifications.

Frequently Asked Questions

Does cfDNA plasma preparation always require two centrifugation steps?

No. The NCI example includes two stages or an appropriate filtration alternative, but individual assays may use a different validated procedure. Follow the approved method for the sample and downstream test.

Can a centrifuge rated at 20,600 ×g perform a 16,000 ×g second spin with any rotor?

No. The exact rotor must reach the required RCF with the actual approved vessel. For example, the YT16R G16-15 has a published 11,840 ×g limit despite the main unit's 20,600 ×g rating.

Why is plasma transfer necessary after the first spin?

It separates the plasma fraction from the blood-cell layer and helps avoid carrying cells into subsequent processing. Simply raising the force on the original whole-blood tube does not reproduce that transfer.

Does the second high-speed spin pellet all cfDNA?

No. In the cited method its purpose is further removal of cellular material and debris, not recovery of all cfDNA as a pellet. Actual recovery and fragment profiles must be evaluated for the chosen protocol.

Is more extracted DNA always a sign of better plasma processing?

No. Greater total DNA may include genomic DNA released from blood cells. Fragment profiles, contamination indicators and downstream analytical performance are more informative than total yield alone.

Is the largest high-speed rotor automatically best for cfDNA?

No. A 12 × 15 mL rotor may have a lower maximum RCF than a smaller-capacity rotor. Match the required force, approved vessel, usable fill volume and batch size together.

Match the Complete cfDNA Sample Pathway, Not a Maximum RCF Label

Effective cfDNA plasma preparation starts with the biological source of contamination and the intended assay. The first spin, transfer step and subsequent treatment have different objectives and potentially different vessel, rotor and handling requirements.

For purchasing, the correct sequence is sample and method → original collection tube → first-stage RCF and rotor → plasma transfer vessel → second-stage RCF and rotor → real batch capacity → complete equipment configuration. This prevents both under-specified high-force rotors and unnecessary equipment choices driven by headline speed alone.

Technical References

US National Cancer Institute (NCI). Cell-Free DNA: Biospecimen Collection and Processing, Biospecimen Evidence-Based Practices. Original release: January 30, 2020; February 5, 2025 revision for accessibility compliance.

International Organization for Standardization. ISO 20186-3:2019, Molecular in vitro diagnostic examinations — Specifications for pre-examination processes for venous whole blood — Part 3: Isolated circulating cell free DNA from plasma.

Clinical Cancer Research. Harmonizing Cell-Free DNA Collection and Processing Practices through Evidence-Based Guidance , 2020.

GlanLab. YT5AR Blood Station Refrigerated Centrifuge and YT16R Desktop High-speed Refrigerated Centrifuge: published product and matched-rotor specifications.

Reference conditions and equipment ratings do not replace the validated assay procedure, collection-vessel instructions or final rotor configuration confirmation.

Need a Centrifuge Configuration for Two-Stage Plasma Processing?

Send the original collection tube, transfer vessel, RCF and time for each stage, expected plasma volume, sample count and destination country. GlanLab can review a suitable machine-and-rotor configuration.

Discuss Your Configuration

GlanLab, with over 20 years of experience, manufactures a full range of centrifuge machines, including benchtop, high-speed, floor-standing, and specialized models in China. We offer distribution, wholesale, OEM services, and single-unit orders at competitive prices. With complete quality certifications and robust after-sales support, GlanLab is your trusted partner for centrifuge supplies.
Contact Us
  +86-18362053005
   inquiry@glanlab.com
  No. 151, Building 60, Houhu Art Park Area D, Yuelu Dist, Changsha, Hunan, China
Leave a Message
Contact Us
Copyright© 2025 Changsha Glanlab Tech Co., Ltd.  Support By jingdian    Sitemap    Privacy Policy